A Wear Formulation of Total Hip Prosthesis for Salat Activity


(*) Corresponding author


Authors' affiliations


DOI: https://doi.org/10.15866/ireme.v13i1.16340

Abstract


The wear on the acetabular liner surface of total hip prosthesis due to human activity cannot be avoided. Salat is an obligation religious activity for Muslims performed five times daily, without exception for patients who used the total hip prosthesis. Salat consists of standing, bowing, prostrating and sitting on the leg motions. Many researchers have investigated the wear on the liner surface for daily activity but not for salat activity. This study has aimed to calculate wear on the acetabular liner surface due to salat activity. A new linear wear model has been proposed in order to estimate the linear wear or the wear depth on the acetabular liner surface. In addition, a volumetric wear model or volume loss on the acetabular liner surface has also been developed based on the new linear wear model. These wear model are used to calculate wear on the acetabular liner surface due to walking and salat activities. Results show that the linear wear due to salat activity is lower than walking activity, because of the intensity of salat activity is lower than walking. However, the combinations of both wear due to walking activity and salat activity need to be considered.
Copyright © 2019 Praise Worthy Prize - All rights reserved.

Keywords


Gait Cycle; Linear Wear Model; Salat Activity; Total Hip Prosthesis; Volumetric Wear

Full Text:

PDF


References


Pew Research Center, The future of world religions: population growth projections, 2010-2050. Why Muslims Are Rising Fastest and the Unaffiliated Are Shrinking as a share of the World’s Population, April 2, 2015, www.pewresearch.org.

The Holy Quran (chapter 47, verse 19; chapter 24, verse 56; chapter 2, verse 110; chapter 2, verse 183; chapter 3, verse 97).

C. V. Sikes, L. P. Lai, M. Schreiber, et al., Instability after total hip arthroplasty treatment with large femoral heads vs constrained liners, The Journal of Arthroplasty, Vol. 23, pp. 59-63, 2008.
https://doi.org/10.1016/j.arth.2008.06.032

J. M. Cuckler, The dislocated total hip: the dreaded 3 am phone call, Seminar in Arthroplasty, Vol. 22, pp. 98-99, 2011.
https://doi.org/10.1053/j.sart.2011.03.012

M. T. Hummel, A. L. Malkani, M. R. Yakkanti, et al., Decreased dislocation after revision total hip arthroplasty using larger femoral head size and posterior capsular repair, The Journal of Arthroplasty, Vol. 24, pp. 73-76, 2009.
https://doi.org/10.1016/j.arth.2009.04.026

J. F. Archard, Contact and rubbing of flat surfaces, Journal of Applied Physics, Vol. 24, pp. 981-988, 1953.
https://doi.org/10.1063/1.1721448

J. P. Paul, Force actions transmitted by joints in the human body, Proceedings of Royal Society of London B, Vol. 192, pp. 163-172, 1976.

O. Calonius, V. Saikko, Analysis of relative motion between femoral head and acetabular and advances in computation of the wear factor for the prosthetic hip joint, Acta Polytechnica, Vol. 43, pp. 43-54, 2003.

D. Dowson, B. Jobbins, A. Seyed-Harraf, An evaluation of the penetration of ceramic femoral heads into polythene acetabular cups, Wear, Vol. 162-164, Part B, pp. 880-889, 1993.
https://doi.org/10.1016/0043-1648(93)90090-9

J. J. Kauzlarich, J. A. Williams, Archard wear and component geometry, Proceedings of The Institution of Mechanical Engineers, Vol. 215, pp. 387-403, 2001.

J. S. S. Wu, J. P. Hung, C. S. Shu, et al., The computer simulation of wear behavior appearing in total hip prosthesis, Computer Methods and Programs in Biomedicine, Vol. 70, n.1, pp. 81-91, 2003.
https://doi.org/10.1016/s0169-2607(01)00199-7

G. K. Sfantos, M. H. Aliabadi, Total hip arthroplasty wear simulation using the boundary element method, Journal of Biomechanics, Vol. 40, pp. 378-389, 2007.
https://doi.org/10.1016/j.jbiomech.2005.12.015

F. Liu, A. Galvin, Z. Jin, et al., A new formulation for the prediction of polyethylene wear in artificial hip joints, Proceedings of The Institution of Mechanical Engineers Part H, Vol. 225, n. 1, pp. 16-24, 2011.
https://doi.org/10.1243/09544119jeim819

F. Liu, J. Fisher, Z. Jin, Effect of motion inputs on the wear prediction of artificial hip joints, Tribology International, Vol. 63, pp. 105-114, 2013.
https://doi.org/10.1016/j.triboint.2012.05.029

M. Alvarez-Vera, J. A. Ortega-Saenz, M. A. L. Hernandez-Rodrigues, A study of the wear performance in a hip simulator of a metal-metal Co-Cr alloy with different boron additions, Wear, Vol. 301, pp. 175-181, 2013.
https://doi.org/10.1016/j.wear.2013.01.085

M. S. Uddin, L. C. Zhang, Predicting the wear of hard-on-hard hip joint prostheses, Wear, Vol. 301, pp. 192-200, 2013.
https://doi.org/10.1016/j.wear.2013.01.009

M. S. Uddin, Wear Measurement and Assessment of Explanted Cross-Linked PE Acetabular Cups Using a CMM, Tribology Transactions, Vol. 57, n. 5, pp. 767-777, 2014.
https://doi.org/10.1080/10402004.2014.911398

S. Shankar, K. Gowthaman, M. S. Uddin, Predicting long-term wear performance of hard-on-hard bearing couples: effect of cup orientation, Medical and Biological Engineering and Computing, Vol. 54, n. 10, pp. 1541-1552, 2015.
https://doi.org/10.1007/s11517-015-1440-2

M. S. Uddin, C. Y. E. Mak, S. A. Callary, Evaluating hip implant wear measurements by CMM technique, Wear, Vol. 364-365, pp. 193-200, 2016.
https://doi.org/10.1016/j.wear.2016.07.017

P. Damm, J. Dymke, R. Ackermann, et al., Friction in total hip joint prosthesis measured in vivo during walking, PLoS ONE 2013., Vol. 8, n. 11, pp. 1-8, 2013.
https://doi.org/10.1371/journal.pone.0078373

G. Bergmann, F. Graichen, A. Rohlmann, et al., Realistic loads for testing hip implants, Bio-medical Material and Engineering, Vol. 20, pp. 65-75, 2010.

G. Bergmann, F. Graichen, A. Rohlmann, Hip joint contact forces during stumbling, Langenbeck’s Archives of Surgery, Vol. 389, n. 1, pp. 53-9, 2004.
https://doi.org/10.1007/s00423-003-0434-y

G. Bergmann, G. Deuretzbacher, M. Heller, et al., Hip contact forces and gait patterns from routine activities, Journal of Biomechanics, Vol. 34, pp. 859-71, 2001.
https://doi.org/10.1016/s0021-9290(01)00040-9

G. Bergmann, F. Graichen, A. Rohlmann, Loads acting at the hip joint in: Sedel, Cabanela (eds) Hip surgery - new materials and developments. M Dunitz, London, pp. 1-8, 1998.

G. Bergmann, F. Graichen, A. Rohlmann, Hip Joint Loading Walking and Running, Measured in Two Patients, Journal of Biomechanics, Vol. 26, n. 8, pp. 969-990, 1993.
https://doi.org/10.1016/0021-9290(93)90058-m

G. Bergmann, F. Graichen, A. Rohlmann, et al., Hip joint forces during load carrying, Clinical Orthopaedics and Related Research, Vol. 335, pp. 190-201, 1997.
https://doi.org/10.1097/00003086-199702000-00018

A. Y. B. Hashim, N. A. A. Osman, W. A. B. W. Abas, et al., Analysis of shalat gait, International Journal of Open Problems in Computer Science and Mathematics, Vol. 3, pp. 552-562, 2010.

J. Jamari, I. B. Anwar, E. Saputra, et al., Range of motion simulation of hip joint movement during salat activity, The Journal of Arthroplasty., Vol. 32, n. 9, pp. 2898-2904, 2017.
https://doi.org/10.1016/j.arth.2017.03.056

E. Saputra, I. B. Anwar, J. Jamari, et al., Finite element analysis of artificial hip joint movement during human activities, Procedia Engineering, Vol. 68, pp. 102-108, 2013.
https://doi.org/10.1016/j.proeng.2013.12.154

J. Jamari, R. Ismail, E. Saputra, et al., The effect of repeated impingement on UHMWPE material in artificial hip joint during salat activities, Advanced Material Research, Vol. 896, pp. 272-275, 2014.
https://doi.org/10.4028/www.scientific.net/amr.896.272

V. Saikko, A three-axis hip joint simulator for wear and friction studies on total hip prosthesis, Proceedings of The Institution of Mechanical Engineers Part H, Vol. 210, pp. 175-185, 1996.
https://doi.org/10.1243/pime_proc_1996_210_410_02

V. Saikko, T. Ahlroos, Type of motion and lubricant in wear simulation of polyethylene acetabular cup, Proceedings of The Institution of Mechanical Engineers Part H, Vol. 213, pp. 301-310, 1999.
https://doi.org/10.1243/0954411991535130

H. A. Mckellop, I. C. Clarke, Degradation and wear of ultra-high-molecular-weight polyethylene. In: Corrosion and degradation of implant materials: second symposium, ASTM STP 859. (Eds. Fraker A.C, Griffin C.D.), Philadelphia: American Society for Testing and Materials, pp. 351-368, 1985.
https://doi.org/10.1520/stp33263s

A. Wang, V. K. Polineni, A. Essner, et al., The significance of nonlinear motion in the wear screening of orthopaedic implant materials, Journal of Testing and Evaluation, Vol. 25, pp. 239-245, 1997.
https://doi.org/10.1520/jte11485j

P. A. Revell, N. Al-Saffar, A. Kobayashi, Biological reaction to debris in relation to joint prosthesis, Proceedings of The Institution of Mechanical Engineers Part H, Vol. 211, pp. 187-197, 1997.
https://doi.org/10.1243/0954411971534304

M. Ungethum, Tribological biomechanical examinations for total joint replacement of the human hip, Ph.D. dissertation, Rheinisch-Wesfalische Technische Hochschule Aachen, 1976.

A. Walter, Investigations on the wear couple biolox forte and earlier alumina materials. In: Performance of the wear couple biolox forte in hip arthroplasty (E. Puhl, W.), Stuttgart, Ferdinand Enke Verlag, pp. 123-135, 1997.

D. Dowson, B. Jobbins, Design and development of a versatile hip joint simulator and a pleminary assessment of wear and creep in Charnley total replacement hip joints, Engineering in Medicine, Vol. 17, pp. 111-117, 1988.
https://doi.org/10.1243/emed_jour_1988_017_031_02

ISO/DIS 14242-1 Draft International Standard. Implants for surgery-Wear of total hip joint prostheses-Part I: Loading and displacement parameters for wear-testing machines and corresponding environmental conditions for test, 2001.

S. L. Smith, A. Unsworth, A five-station hip joint simulator. Proceedings of The Institution of Mechanical Engineers Part H, Vol. 215, pp. 61-64, 2001.
https://doi.org/10.1243/0954411011533535

P. S. M. Barbour, M. H. Stone, J. Fisher, A hip joint simulator study using simplified loading and motion cycles generating physiological wear paths and rates, Proceedings of The Institution of Mechanical Engineers Part H, Vol. 213, pp. 455-467, 1999.
https://doi.org/10.1243/0954411991535068

A. A. J. Goldsmith, D. Dowson, A multi-station hip joint simulator study of the performance of 22 mm diameter zirconia-ultra-high molecular polyethilene total replacement hip joints. Proceedings of The Institution of Mechanical Engineers Part H, Vol. 213, pp. 77-90, 1999.
https://doi.org/10.1243/0954411991534825

R. C. Johnston, G. L. Smidt, Measurement of hip-joint motion during walking-evaluation of an electro goniometric method, The Journal of Bone & Joint Surgery, Vol. 51, n. 6, pp. 1083-1094, 1969.
https://doi.org/10.2106/00004623-196951060-00003

A. van Beek, Advanced engineering design: lifetime performance and reliability, (Delft University of Technology, 2012).

B. B. Seedhom, N. C. Wallbridge, Walking activities and wear of prostheses. Annals of The Rheumatic Diseases, Vol. 44, pp. 838-843, 1985.
https://doi.org/10.1136/ard.44.12.838

T. P. Schmalzried, E. S. Szuszczewicz, M. R. Northfield, et al., Quantitative assessment of walking activity after total hip or knee replacement, The Journal of Bone & Joint Surgery American Volume, Vol. 80, pp. 4-9.
https://doi.org/10.2106/00004623-199801000-00010


Refbacks

  • There are currently no refbacks.



Please send any question about this web site to info@praiseworthyprize.com
Copyright © 2005-2024 Praise Worthy Prize